OAR 333-116-0410: Availability of Survey Instrument
Edition: Current OARD snapshotEffective: 2018-02-01Last verified: 2026-09-07
A licensee authorized to use radioactive material as a sealed source for diagnostic purposes must have available for use a portable radiation detection survey instrument capable of detecting dose rates over the range one uSv (0.1 mrem) per hour to 100 mrem (one mSv) per hour and a portable radiation measurement survey instrument capable of measuring dose rates over the range ten uSv (1 mrem) per hour to ten mSv (1000 mrem) per hour. The instrument must be operable and calibrated in accordance with OAR 333-116-0170.
Official history and status
Statutory/Other Authority: ORS 453.635
Statutes/Other Implemented: ORS 453.605 - 453.807
History: PH 29-2018, minor correction filed 02/01/2018, effective 02/01/2018 PH 4-2007, f. & cert. ef. 3-1-07 PH 12-2006, f. & cert. ef. 6-16-06 PH 36-2004, f. & cert. ef. 12-1-04 PH 31-2004(Temp), f. & cert. ef. 10-8-04 thru 4-5-05 PH 3-2003, f. & cert. ef. 3-27-03 HD 1-1991, f. & cert. ef. 1-8-91