OAR 333-106-0320: Additional Requirements for Radiographic Machines: Radiation from Capacitor Energy Storage Equipment in Standby Status
Edition: Current OARD snapshotEffective: 2010-02-16Last verified: 2026-09-07
Radiation emitted from the X-ray tube when the exposure switch or timer is not activated shall not exceed a rate of two milliroentgens (0.516 uC/kg) per hour at five centimeters from any accessible surface of the diagnostic source assembly, with the beam-limiting device fully open.
Official history and status
Statutory/Other Authority: ORS 453.605 - 453.755
Statutes/Other Implemented: ORS 453.605 - 453.755
History: PH 4-2010, f. & cert. ef. 2-16-10 PH 14-2008, f. & cert. ef. 9-15-08 HD 1-1991, f. & cert. ef. 1-8-91 HD 4-1985, f. & ef. 3-20-85