OAR 333-106-0230: Fluoroscopic X-ray Systems Requirements: Fluoroscopic Timer
Edition: Current OARD snapshotEffective: 2008-09-15Last verified: 2026-09-07
(1) Means shall be provided to present the cumulative on-time of the fluoroscopic tube. (2) The maximum cumulative time of the timing device shall not exceed five minutes without resetting. (3) A signal audible to the fluoroscopist, or the appropriate operator, shall indicate the completion of any preset cumulative on-time; or if no audible signal is provided, the exposure shall terminate.
Official history and status
Statutory/Other Authority: ORS 453.752 - 453.775
Statutes/Other Implemented: ORS 453.752 - 453.775
History: PH 14-2008, f. & cert. ef. 9-15-08 HD 4-1985, f. & ef. 3-20-85